DESCRIPTION
- Single channel EMG Triggered stimulation
- Portable desk top device supplied with a desk top stand
- Passive and Active(EMG and Stimulation)
- Pre-set clinical modes for: Incontinence, sports and rehabilitation.
- Custom clinical modes: EMG, STIM and ETS
- Locks programmes and records patient home compliance on a daily basis.
- Portable light and user friendly
New and Improved touch-screen with full colour display. The new user friendly interface makes it a self-explanatory unit with the interactive learning on the go, including electrode placement, running EMG graph as well as many other helpful tools.
Pre-set programmes for Incontinence treatment and neuro-muscular Rehabilitation.
Designed for Physiotherapist in mind to be used as a portable tool. Suitable as a personal trainer for in-clinic or home use (under guidance).
Single channel EMG (Biofeedback) combined with 1 channels of NMES (stimulation).
Single channel ETS (EMG triggered stimulation) with single channel stimulation.
Simple single-phase or multiphase operation: multiphase allows EMG/ETS/STIM to be combined in one programme.
Multilingual LCD screen and voice prompts for Biofeedback.
Manual and Automatic threshold control.
EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
Stand alone device or used in conjunction with established multilingual software.
Now with Bluetooth connectivity! The unit sends up to 10 metres of wireless data to PC software.
For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
4 x AAA batteries {No mains required}
1. EMG
1.1 EMG Range: 0.2 to 2000 μV RMS (continuous)
1.2 Sensitivity: 0.l μV RMS
1.3 Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
1.4 Selectable Bandpass filter – 3db Bandwidth,
a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% – Reading
below 235 microvolts
10 Hz +/-3 Hz to 370 Hz +/- 10% – Reading
above 235 microvolts
b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
1.5 Notch filter: 50 Hz (Canada 60Hz) – 33 dbs (0. 1% accuracy)
1.6 Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
1.7 Work / Rest periods: 2-99 seconds
1.8 Number of Trials: 2-99
2. NMES, TENS and ETS
2.1 Single channel Stimulator
2.2 Amplitude: 0-90 mA into 500 Ohm load – actual mA will tend to be less
than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes
in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms
load the maximum will be limited to 50 mA.
2.3 Type: Constant current, maximum output
voltage 70 Volts +5 / -10 Volts
2.4 Waveform: Symmetrical, rectangular,
bi-phasic with net zero DC current
2.5 Pulse width selection: 50-330 μS (10% accuracy)
2.6 Pulse rate selection: 2-100 Hz (5% accuracy)
2.7 Work / Rest periods: 2-99 seconds
2.8 Time 1 – 99 minutes
2.9 Ramp up time: 0.1 – 9.9 seconds
2.10 Preset and user programmable treatment Programs
2.11 Automatic output shut off with detection of open electrode above 10 mA
(+/- 2mA).